TechFlo Products

Bench and Desktop Fume Cabinets

Contain and filter for a cleaner, healthier working environment within dental labs, electronic design and rework, pharmaceutical labs, design and technology, science in education and more . . . .

For use in numerous situations where external venting is impractical but where contaminant capture and filtration is necessary.

Applications include the filtration of fumes, vapour and particulates emanating from:

Monomers, adhesives, hazardous powders/particulates, conformal coatings, resins, sealants, solders, light grinding particles and dusts, organic solvents etc.

T500S-C and T700S-C

Air Filtration Cabinets

Download Printable PDF Specs of the TechFlo Model T500S-C and T700S-C

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Techflo T500S-C 700S-C  air filtration cabinets

TechFlo Air Filtration Units

Model No's T500S-C and 700S-C

NOTE: These units are recirculatory; i.e. filtered air is exhausted back into the workplace.

Description

Fan/filter casing (steel construction finished in dove grey) - clear PETG folding plastic fume shield - particulate (Intake Filter) - bonded carbon filter - centrifugal blower(s) (T700-C has 2 Blowers) , external rotor motor type brushless, UL, VDE and CSA approved - sealed electrical wiring and switch, downstream of filters. Note: T700S-C has a dual airspeed function to facilitate higher vapour adsorption.

Operation

All WS filtration and extraction systems carry a lifetime warranty, are quiet in operation and can be assembled in seconds. Place onto a desk, bench or trolley and then simply plug in and switch on. The folding PETG shield can be stored if not required thus reducing the unit footprint considerably. Handling and / or use of potentially harmful substances is carried out in front of the intake filter area. Cleaned air is vented to the top rear of the units

'A' Dimensions = T500S-C (Weight 25kgs) 'B' Dimensions = T700S-C (Weight 37kgs)

 

model T460 and T670 Dimensions

Filtration

The particulate (intake) filter is a three stage graduated fibre filter (filtration to 5.0 micron particles) with a high 'dust' holding capacity for long life. (Optional HEPA filters available to 0.3micron particles). This is followed by a high grade bonded carbon filter. (3.5kg T500 and 7kg T700)

Through the process of filtration, adsorption and dilution, operator exposure to potentially hazardous substances is vastly reduced or eliminated

Specification

Model
No.
Fan Motor  *
Spec.
Air Volume
at Free Air
Ave. Air Velocity
at Filter Face
Ave. Air Velocity
at Shield Face
Noise Level
LA eq dB(A)
T500S-C 230V 50Hz 105W 355 m3/hr 0.50m/sec 0.38 m/sec     60
T700S-C 230V 50Hz 210W 710 m3/hr 0.57m/sec 0.55 m/sec ** 62

* 110V 60Hz available if required.
** T700S-C at slow setting, air velocity is 0.32 m/sec at hood face

These units are designed for immediate operator protection from concentrations of airborne contaminants whereby all gases, fumes, vapours and particulates at the front of the unit are drawn into the intake filter. Cleaned air is exhausted to the top rear of the cabinet. See TechFlo (T660's and T930's) Fume Hoods and Cabinets for higher specification requirements.

CoSHH requirementsDemonstrating adequate control of the risks to the health of employees students and the general public, from dusts, powders, mists, fumes and vapours is the fundemental requirement of the Control of Substances Hazardous to Health (CoSHH).

Download Printable PDF Specs of the TechFlo Model T500S-C/T700S-C

 

Please contact WS for advice on the choice of machine for any given situation.
 
We go to great lengths to ensure accuracy in our literature, however we cannot be held responsible for errors or omissions. Information is given for guidance only and we reserve the right to alter specification.